INSTRUCTION FOR USE
MULTIBAND LIGATOR AND SET
FOCUZ BAND™
Doc No.: FMT/IFU/MLSEN, Rev.1.0, 01-05-2026
The Multiband Ligator (FOCUZ BAND TM) is used via endoscope to legate esophageal varies at and above the esophago-gastricl junction. The Multiband Ligator and Set is supplied non-sterile and is intended for single use only.
The contraindications include without being limited to cricopharyngeal or esophageal narrowing/stricture, tortuous esophageal diverticulitis, known or suspected esophageal perforation, asymptomatic rings or webs, coagulopathy. Use of ligation bands is contraindicated in patients with a known hyper sensitivity to latex.
Possible complications associated with the gastrointestinal endoscopy include but not limited to perforations, fever, aspirations, hemorrhage, infections, hypotension, allergic reactions to medication, cardiac arrhythmia or arrest, and respiratory depression or arrest. Furthermore, complications that may occur with esophageal banding include but not limited to chest tightness, retrosternal pain, dysphasia or odynophagia, post-EVL ulcer with or without bleeding, increase of portal hypertensive gastropathy, bacteremia, nausea, laryngeal, retrosternal pain, laceration, stricture formation, esophageal perforation and obstructions.
It is vital that biopsy channel size of the endoscope is co-ordinate with compatible devices in order to obtain the optimal results during the procedure. The minimum biopsy channel diameter required is 2.8 mm and length of up to 150 cm. The outer diameter of the scope to be used must be from 8.6 mm up to 10.8 mm. Band ligation may not be effective when applied to small varices. It is vital to ensure the understanding of the technical principles, clinical applications and associated risks with endoscopy as well as endoscopic banding before using this product. This product must only be used by or under direct supervision of physician thoroughly trained in therapeutic endoscopy as well as variceal banding. Esophageal ligators are not intended for ligation of varices below the gastro-esophageal junction. Banding should begin at the gastro-esophageal junction and proceed up the esophagus in spiral manner, respecting a minimum of 2cm in between two band ligations. Passing the endoscope over an already legated varix may dislodge the FOCUZ BAND™ from its position. Routine endoscopic examination by the physician is recommended to confirm the diagnosis requiring treatment of esophageal varices or internal hemorrhoids prior to assembling the Multiband Ligator and Set.

This device is intended for use in adult patients (≥18 years of age).
This device is intended for use by trained and qualified healthcare professionals experienced in endoscopic and/or interventional procedures.
After use, this product may be a potential biohazard. Handle and dispose of in accordance with hospital, local and administrative laws and regulations.
In case of questions or difficulties concerning our instruments please Contact your local distributor or Focuz Medical Technologies LLP directly during Regular working hours.